DC Characteristics

Accuracy

  • ±(% of reading + % of range)
  • The specifications are for the following conditions: 6 1/2 digit resolution, minimum two-hour warm up, and auto trigger mode.
  • Use the null function for the 2-wire / 4-wire resistance measurement method.

DC voltage

RangeResolutionInput resistance1 year (23 °C ± 5 °C)
100.0000 mV0.1 μV> 10 GΩ0.0050+0.0035
1.000000 V1.0 μV> 10 GΩ0.0040+0.0007
10.00000 V10 μV> 10 GΩ0.0035+0.0005
100.0000 V100 μV10 MΩ0.0045+0.0006
1000.000 V1 mV10 MΩ0.0045+0.0010

DC current

RangeResolutionInput resistance1 year (23 °C ± 5 °C)
10.00000 mA10 nA5.1 Ω0.050+0.020
100.0000 mA100 nA5.1 Ω0.050+0.005
1.000000 A1 μA0.1 Ω0.100+0.010
3.000000 A10 μA0.1 Ω0.120+0.020

Resistance

RangeResolutionInput resistance1 year (23 °C ± 5 °C)
100.0000 Ω100 μΩ1 mA0.010+0.004
1.000000 kΩ1 mΩ1 mA0.010+0.001
10.00000 kΩ10 mΩ100 μA0.010+0.001
100.0000 kΩ100 mΩ10 μA0.010+0.001
1.000000 MΩ1 Ω5 μA0.010+0.001
10.00000 MΩ10 Ω500 nA0.040+0.001
100.0000 MΩ100 Ω500 nA||10 MΩ0.800+0.010

Diode test

RangeResolutionTest current1 year (23 °C ± 5 °C)
1.0000 V10 μV1 mA0.010+0.020

Continuity test

RangeResolutionTest current1 year (23 °C ± 5 °C)
1 kΩ10 mΩ1 mA0.010+0.030

Measurement characteristics

ItemSpecifications
DC voltage measurement : OverrangePermits voltages that are up to 20 % over the range except when the 1000 V range is in use
DC voltage measurement : Input bias currentLess than 30 pA (at 25 °C)
DC voltage measurement : Input voltage protection1000 V for all ranges
DC current measurement : OverrangePermits currents that are up to 20 % over the range except when the 3 A range is in use
Resistance measurement : Maximum test lead resistance10 Ω (100 Ω range)
100 Ω (1 kΩ range)
1 kΩ (other ranges)
Resistance measurement : Input voltage protection1000 V for all ranges