TOS5200_Comparison with the Kikusui previous model
| Specification comparison item | TOS5200 | TOS5050A |
|---|
| Appearance |  |  |
|---|
| Test voltage | Output method | PWM switching amp system | Slide transformer method |
|---|
| Distortion | 3 % or less | As per commercial power supply waveform |
|---|
| Frequency | 50 Hz or 60 Hz | Synchronized with commercial power supply waveform |
|---|
| Output voltage waveform | Sine | Commercial power supply waveform |
|---|
| Voltage regulation | 10 % or less | 15 % or less |
|---|
| Input voltage variation | ±0.3 % | – |
|---|
| Test time | Minimum setting value | 0.1 s | 0.5 s |
|---|
| Accuracy | ±(100 ppm + 20 ms) excluding Fall Time | ±(100 ppm + 20 ms) |
|---|
| Judgment feature | Upper limit setting | AC: 0.01 mA to 110 mA | AC: 0.1 mA to 110 mA |
|---|
| Lower limit setting | AC: 0.01 mA to 110 mA | AC: 0.1 mA to 110 mA |
|---|
| Accuracy | 1.00 mA ≤ i: ±(1.5 % of set),
i <1.00 mA: ±(1.5 % of set +30 μA) | Upper limit: ±(5 % + 20 μA) |
|---|
| Rise Time and Fall Time control features | 3 | – |
|---|
| Voltmeter | Display | Digital | Digital, analog |
|---|
| Measurement accuracy | ±1.5 % of reading (V>500V) | ±1.5 % f.s |
|---|
| Voltmeter/Ammeter | Measurement method | True rms/mean-value | Mean-value/rms value |
|---|
| RS232C Interface | POWER Switch and all function except Key lock | Output for data, test result |
|---|
| Weight | Approx. 14 kg (30.9 lbs) | Approx. 15 kg (33.07 lbs) |
|---|
| Other | Input power supply | 100 Vac to 240 Vac | 100 V ± 10 % |
|---|