| Power requirement | Nominal voltage range (Allowable voltage range) | 100 Vac to 240 Vac. (90 Vac to 250 Vac). |
|---|---|---|
| Power consumption | 30 VA or less. | |
| Nominal frequency range (Allowable frequency range) | 50 Hz to 60 Hz. (45 Hz to 65 Hz). | |
| Insulation resistance | Between AC line and chassis | 100 M or more at 500 Vdc. |
| Withstanding voltage | Between AC line and chassis | 1 500 Vac for 1 minute. |
| Earth continuity | 25 Aac/0.1 Ω or less | |
| Electromagnetic Compatibility (EMC)*2 | Complies with the requirements of the standard below. EN 61326 Applicable condition All of the wires and wires connected to the KSG3600 are less than 3 m in length. | |
| Safety *2 | Complies with the requirements of the standard below. EN 61010-1 Class I Pollution Degree2 | |
| Environment | Operating environment | Indoor use, Overvoltage Category II. |
| Operating temperature and humidity range | 0 C to 40 C, 0 % rh to 80 %rh (no condensation). | |
| Temperature and humidity range for guaranteed specifications | 23 C±5 C, 0 % rh to 80 %rh (no condensation). | |
| Storage temperature and humidity range | -20 C to 70 C, 0 % rh to 90 %rh (no condensation). | |
| Altitude | Up to 2000 m | |
| Dimensions | 280(285) W × 44(60) H × 270(330)D mm | |
| Weight | Approx. 3 kg (6.61 lb) | |
| Accessories | Power cord | SVT3 18AWG: 1 pc. With plug Cord length: Approx. 2.5 m. |
| Operation Manuals | 1 copy | |
| BNC-banana plug cable | 1 pc. Cord length: 1 m. | |
| Spare fuse *3 | 1 pc | |
*2 Not applicable to custom order models.*3 Provided inside the KSG3600.