AC measurement characteristics

ItemDME1700DME1701
Common specificationsAC measurement modeWith anti-aliasing filter AC-coupled digital sampling
Crest factor
(except sine waves)
Crest factor selectionThe larger of the following crest factors:
• Up to 3:1 of full scale input • Max. 10:1
Measurement range selection in auto rangeOptimal measurement range for the crest factor up to 10:1
Applying the accuracy characteristics[Crest factor] × [fundamental frequency] ≤ 3 kHz
(applied to all crest factors)
Volt-Hertz product8 × 107 V*Hz or below
Common mode rejection ratio (If LO terminal lead imbalance is 1 kΩ)>>70 dB
AC filterMeasuring input signals with frequencies greater than the set AC filter bandwidth
Effect of AC filter bandwidth on aperture time• 2 ms at 200 Hz
• 20 ms at 20 Hz
• 200 ms at 3 Hz
Input impedance (if parallel connection < 100 pF)1.1 MΩ ± 2 %
Measurement range additional errorIf the input is below 50 kHz and is 1 % to 5 % of the measurement rangeAdding 0.1 % of the measurement range to the accuracy
If the input is 50 kHz to 100 kHzAdding 0.13 % of the measurement range to the accuracy
AC voltage measurement
(ACV)
Over-rangeFor 100 mV to 100 V ranges20 % over-range allowed
For 750 V rangeNo over-range
Maximum DC voltageAll ACV measurement ranges400 V
AC current measurement
(ACI)
Over-rangeFor 10 mA to 1 A ranges20 % over-range allowed
For 3 A rangeNo over-range