Output section
| Output voltage range | -25 V to -1000 V | |
|---|---|---|
| Resolution | 1V | |
| Setting accuracy | ±(1.5% of setting + 2V) | |
| Max. rated load | 1W(−1000Vdc/1mA) | |
| Max. rated current | 1mA | |
| Output terminal | Output type | Floating |
| Isolation voltage | ±1000 VDC | |
| Ripple | 1 kV no load | 2 Vp-p or less |
| Max. rated load | 10 Vp-p or less | |
| Short-circuit current | 12 mA or less | |
| Output rise time | 50 ms or less (10 % to 90 %) [no load] | |
| Discharge feature | Forced discharge at the end of test (discharge resistance: 25 kΩ) | |
Voltmeter
| Measurement range | 0 V to -1200 V |
|---|---|
| Resolution | 1V |
| Accuracy | ±(1% of reading + 1V) |
Resistance meter
| Measurement range | 0.01 MΩ to 5000 MΩ (In the range of over 100 nA to a maximum rated current of 1 mA) | |
|---|---|---|
| Display | R < 10.0 MΩ:□.□□MΩ 10.0MΩ≦ R < 100.0MΩ:□□.□MΩ 100.0MΩ ≦ R < 1000MΩ:□□□MΩ 1000MΩ ≦ R ≦ 5000MΩ:□□□□MΩ R = measured insulation resistance | |
| Accuracy | 100nA < i ≦ 200nA:±(10% of reading) 200nA < i ≦ 1μA:±(5% of reading) 1μA < i ≦1mA:±(2% of reading) [In the humidity range of 20 %rh to 70 %rh (no condensation), with no disturbance such as swinging of the test leadwire] i =measured output-voltage value/measured resistance value | |
| Measurement range | The current measurement range is selectable between AUTO and FIX. | |
| AUTO | Automatically changes the current measurement range according to the measured current value. | |
| FIX | Fixes the current measurement range based on the output voltage set value and LOWER set value (in UPPER OFF status). | |
| Holding function | The resistance measured at the end of test is held during the PASS interval | |
Judgment feature
| Judgment method and judgment operation |
| ||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Setting range for the upper reference value (UPPER) | 0.01 MΩ to 5000 MΩ [In the range of the maximum rated current or less] | ||||||||||||||||||||||||||||||||||||||
| Setting range for the lower reference value (LOWER) | 0.01 MΩ to 5000 MΩ [In the range of the maximum rated current or less] | ||||||||||||||||||||||||||||||||||||||
| Judgment accuracy For both UPPER and LOWER |
The lower judgment requires a test duration of 0.5 s or more after the wait time has expired. It also requires a wait time of 1.0 s or more for a lower judgment of 200 nA or less.] | ||||||||||||||||||||||||||||||||||||||
Time
| Setting range for the test duration (TEST TIME) | 0.5 s to 999 s (TIMER OFF function provided) |
|---|---|
| Setting range for the wait time (WAIT TIME) | 0.3 s to 10 s [TEST TIME > WAIT TIME] |
| Accuracy | ±(100ppm + 20ms) |