TOS7200_Interface and Other Functions

REMOTE

6-pin mini-DIN connector on the front panel The optional remote controller RC01-TOS or RC02-TOS is connected to remotely control starting/stopping of a test (note that a DIN-mini DIN adapter is required).

SIGNAL I/O D-SUB 25-pin connector on the rear panel

No.Signal nameI/ODescription
1PM0ILSB*1Pin Configuration for the SIGNAL I/O Connector]
2PM1I*1
3PM2I*1
4PM3IMSB*1
5N.C  
6N.C  
7N.C  
8N.C  
9STBIInput terminal for the strobe signal of the panel memory
10N.C  
11N.C  
12N.C  
13COM Circuit common (chassis potential)
14H.V ONOON during a test or while a voltage remains between the output terminals
15TESTOON during a test
16PASSOON for approx. 0.2 seconds when PASS judgment is made, or continuously ON while PASS HOLD is activated
17U FAILOContinuously ON if an insulation resistance equal to or exceed-ing the upper resistance is detected, resulting in FAIL judgment
18L FAILOContinuously ON if an insulation resistance equal to or falling below the lower resistance is detected, resulting in FAIL judg-ment
19READYOON during standby
20N.C  
21STARTIStart signal input terminal
22STOPIStop signal input terminal
23ENABLEIRemote control enable signal input terminal
24N.C  
25COM Circuit common (chassis potential)
Input specificationsHigh-level input voltage11 V to 15 VAll input signals are active Low controlled. The input terminal is pulled up to +12 V using a resistor. Opening the input terminal is equivalent to inputting a high-level signal.
Low-level input voltage0 V to 4 V
Low-level input current-5 mA maximum
Input time width5 ms minimum
Output specificationsOutput methodOpen collector output (4.5 V to 30 V DC)
Output withstanding voltage30 V DC
Output saturation voltageApprox. 1.1 V (at 25°C)
Maximum output current400 mA (TOTAL)

* 1-digit BCD active LOW input. Panel memory’s selection signal input terminal

ANALOG OUT

Outputs a logarithmically compressed voltage corresponding to the measured resistance value

Vo=log(1+Rx/1MΩ)
where Rx = measured resistance value (1 MΩ: 0.30 V; 10 MΩ: 1.04 V; 100 MΩ: 2.00 V; 1000 MΩ: 3.00 V; 10000 MΩ or more: 4.00 V). Output impedance: 1 kΩ
Output impedance: 1 kΩ
COMAnalog output-circuit common
Accuracy±(2 % of full scale)

RS232C

D-SUB 9-pin connector on the rear panel (compliant with EIA-232-D)
All functions other than the POWER switch and KEYLOCK function are remotely controllable

Baud rate9600 bps / 19200 bps / 38400 bps (data: 8 bits; parity: none; stop bit: 2 bits fixed)

Display

7-segment LED, 4-digit voltage display, 4-digit insulation resistance display, and 3-digit time display

Memory function

A maximum of 10 types of test conditions can be stored in memory.

Backup battery life

3 years or more (at 25 °C)

TEST MODE

MOMENTARYA test is conducted only when the START switch is pressed.
FAIL MODEDisables cancellation of FAIL judgment using a stop signal via remote control.
DOUBLE ACTIONStarts a test only when the STOP switch is pressed and the START switch is pressed within approximately a half-second.
PASS HOLDAllows the time of holding PASS judgment to be set to 0.2 s or HOLD.

KEYLOCK

Places the tester in a state in which no keystroke other than the START/STOP switch is accepted.