REMOTE
6-pin mini-DIN connector on the front panel The optional remote controller RC01-TOS or RC02-TOS is connected to remotely control starting/stopping of a test (note that a DIN-mini DIN adapter is required).
SIGNAL I/O D-SUB 25-pin connector on the rear panel
| No. | Signal name | I/O | Description | ||
|---|---|---|---|---|---|
| 1 | PM0 | I | LSB*1 | ![]() | |
| 2 | PM1 | I | *1 | ||
| 3 | PM2 | I | *1 | ||
| 4 | PM3 | I | MSB*1 | ||
| 5 | N.C | ||||
| 6 | N.C | ||||
| 7 | N.C | ||||
| 8 | N.C | ||||
| 9 | STB | I | Input terminal for the strobe signal of the panel memory | ||
| 10 | N.C | ||||
| 11 | N.C | ||||
| 12 | N.C | ||||
| 13 | COM | Circuit common (chassis potential) | |||
| 14 | H.V ON | O | ON during a test or while a voltage remains between the output terminals | ||
| 15 | TEST | O | ON during a test | ||
| 16 | PASS | O | ON for approx. 0.2 seconds when PASS judgment is made, or continuously ON while PASS HOLD is activated | ||
| 17 | U FAIL | O | Continuously ON if an insulation resistance equal to or exceed-ing the upper resistance is detected, resulting in FAIL judgment | ||
| 18 | L FAIL | O | Continuously ON if an insulation resistance equal to or falling below the lower resistance is detected, resulting in FAIL judg-ment | ||
| 19 | READY | O | ON during standby | ||
| 20 | N.C | ||||
| 21 | START | I | Start signal input terminal | ||
| 22 | STOP | I | Stop signal input terminal | ||
| 23 | ENABLE | I | Remote control enable signal input terminal | ||
| 24 | N.C | ||||
| 25 | COM | Circuit common (chassis potential) | |||
| Input specifications | High-level input voltage | 11 V to 15 V | All input signals are active Low controlled. The input terminal is pulled up to +12 V using a resistor. Opening the input terminal is equivalent to inputting a high-level signal. | ||
| Low-level input voltage | 0 V to 4 V | ||||
| Low-level input current | -5 mA maximum | ||||
| Input time width | 5 ms minimum | ||||
| Output specifications | Output method | Open collector output (4.5 V to 30 V DC) | |||
| Output withstanding voltage | 30 V DC | ||||
| Output saturation voltage | Approx. 1.1 V (at 25°C) | ||||
| Maximum output current | 400 mA (TOTAL) | ||||
* 1-digit BCD active LOW input. Panel memory’s selection signal input terminal
ANALOG OUT
Outputs a logarithmically compressed voltage corresponding to the measured resistance value
| + | Vo=log(1+Rx/1MΩ) where Rx = measured resistance value (1 MΩ: 0.30 V; 10 MΩ: 1.04 V; 100 MΩ: 2.00 V; 1000 MΩ: 3.00 V; 10000 MΩ or more: 4.00 V). Output impedance: 1 kΩ Output impedance: 1 kΩ |
|---|---|
| COM | Analog output-circuit common |
| Accuracy | ±(2 % of full scale) |
RS232C
D-SUB 9-pin connector on the rear panel (compliant with EIA-232-D)
All functions other than the POWER switch and KEYLOCK function are remotely controllable
| Baud rate | 9600 bps / 19200 bps / 38400 bps (data: 8 bits; parity: none; stop bit: 2 bits fixed) |
|---|
Display
7-segment LED, 4-digit voltage display, 4-digit insulation resistance display, and 3-digit time display
Memory function
A maximum of 10 types of test conditions can be stored in memory.
Backup battery life
3 years or more (at 25 °C)
TEST MODE
| MOMENTARY | A test is conducted only when the START switch is pressed. |
|---|---|
| FAIL MODE | Disables cancellation of FAIL judgment using a stop signal via remote control. |
| DOUBLE ACTION | Starts a test only when the STOP switch is pressed and the START switch is pressed within approximately a half-second. |
| PASS HOLD | Allows the time of holding PASS judgment to be set to 0.2 s or HOLD. |
KEYLOCK
Places the tester in a state in which no keystroke other than the START/STOP switch is accepted.
![Pin Configuration for the SIGNAL I/O Connector]](https://kikusui.co.jp/kiku/wp-content/uploads/2021/02/TOS7200-table.jpg)